Meter & Probe worked guide
Estimate the effect of a voltmeter on a high impedance source
If a sensor node falls from an expected 5 V to about 4.55 V when you connect a meter, the meter may be part of the explanation. A 10 MΩ input is a load, even when its display has plenty of digits.
Content reviewed · Examples are calculations, not tests of a supplied instrument.
Model the source as a voltage plus a resistance
For a steady DC signal, use the source’s equivalent resistance and the meter input resistance on the exact function and range. In a simple linear model, Vind = Vopen × Rin ÷ (Rsource + Rin). Use the manual for the input resistance; it can depend on function, range or mode. Cable leakage and capacitance can matter elsewhere, especially for changing signals.
Two 1 MΩ resistors across an ideal 10 V supply have a 5 V midpoint and a 0.5 MΩ equivalent source resistance. A 10 MΩ voltmeter lowers the calculated midpoint to about 4.762 V. Use the divider's equivalent resistance, not simply one resistor's value.

Two bench examples with the same meter
With an ideal open-circuit 5.00 V source, a 1.00 MΩ source resistance and a 10.0 MΩ input, the connected voltage is about 4.55 V, roughly 9.1% below unloaded. Reduce the source resistance to 10 kΩ and the same meter sees approximately 4.995 V. The display counts did not change; the interaction with the circuit changed.
A larger input resistance reduces this effect. A low impedance mode deliberately loads a circuit more and must not be treated as a higher accuracy version of the normal mode. Use a suitable input or buffer only within its own verified conditions; do not use this calculation to choose equipment for live electrical testing.
Separate two effects in the measurement record
List the estimated loading and the display error as separate lines. To read a named meter’s numerical specification, see the UT139C DC voltage example.

Sources and review
Reviewed . Numerical examples are calculations or illustrations.