Tool 01 / bench measurement

How do I calculate the error range of a multimeter reading?

Enter the reading and the instrument specification line. The interval combines percentage of reading with the digit contribution for the active range.

Calculate an interval

The example defaults describe a 20 mA reading. Replace them with the exact range in your manual.

Calculating…

What this can establish

Specification interval

Whether the stated tolerance fits around the indication under the manual's reference conditions.

Resolution contribution

How the digit term changes when the active range changes.

Decision boundary

Whether a tighter requirement needs another method or a better specified instrument.

What this cannot establish

A calculation does not verify a shipped sample, lead contact, source loading, temperature, calibration status or a UK installation. Counts describe display capacity; they do not replace the accuracy line.

Try this in a real bench case

For the UT139C 60.00 V range, the existing guide uses the published ±(0.7% + 3 digits) line and a 0.01 V step. That produces a specification-only interval around a 12.00 V example. Read the exact manual for the model and range in hand.

Illustrative multimeter test leads on a bench

Illustrative Wikimedia Commons scene; not an independent Meter & Probe measurement.

Next measurement

If the interval is too wide, record the actual range and digit step, then compare the same source with a method whose loading and bandwidth suit the question. Keep the original reading and conditions in your notes.

Read counts vs accuracy → See the UT139C example →

Questions this template helps answer

  • Do more multimeter counts mean better accuracy?
  • How do I calculate multimeter accuracy from percent and counts?
  • What does ±(percent of reading plus digits) mean?
  • How do I calculate the error range of a multimeter reading?
  • Can my multimeter resolve this current change?
  • Why do two multimeters show different readings?
  • Is display resolution the same as measurement accuracy?

How do I calculate a multimeter accuracy interval?

For a specification stated as ±(% of reading + counts), add the percentage of the absolute displayed reading to the count term. Each count is one smallest displayed step on the active range. Subtract and add the resulting error limit to form the interval.

Worked example: For 20.00 mA, ±(0.5% + 2 counts) and a 0.01 mA step, the limit is ±0.12 mA and the interval is 19.88–20.12 mA.

Assumptions and limits: Use the manufacturer’s range and conditions. The calculation does not include extra effects such as lead resistance, source loading or temperature outside the specified limits.

References: Reading the UT139C accuracy specification · UNI-T UT139 series specification table

Common questions

My meter is on the 200 m range. Is the number millivolts or milliamps?

The m prefix means milli. The dial function and the unit on the display tell you whether it is mV or mA.

How do I turn ±(1% + 3 digits) into an error limit?

Use 1% of the magnitude of the displayed reading and add three times the value of the last display digit on that range. Check the manual for the conditions that make the specification valid.

Is one display digit the same thing as the meter’s accuracy?

One digit gives the display step. Accuracy includes the specified reading percentage, any count term and the conditions for that range.

Free online measurement tools

Free multimeter accuracy calculator

Calculate a specification-only error interval from a reading, percentage, counts and display step. Use this free online measurement tool without creating an account. The worked example and limits on this page explain how to read the result.

No sign-up. No subscription. All eight tools are free to use.

Browse all free multimeter tools →

Practical questions

Questions about using the error-interval calculator

How do I enter a multimeter accuracy of 0.5% plus 2 digits?

Enter 0.5 for the percentage, 2 for the count term and the active range’s smallest display step. The step uses the same unit as the reading.

Link to this answer
What is the error interval for 10.00 V at 0.5% plus 2 counts?

If the active step is 0.01 V, the limit is 0.05 V plus 0.02 V. The specification-only interval is 9.93 V to 10.07 V.

Link to this answer
Should the percentage accuracy input be 0.5 or 0.005?

Enter 0.5 for a stated 0.5%. The calculator converts the percentage internally; entering 0.005 would describe a different specification.

Link to this answer
What does the digit-step box mean in a multimeter error calculator?

Enter the value represented by one count on the active range, such as 0.01 V. Do not enter the total display capacity.

Link to this answer
Can this calculator handle a negative voltage indication?

The percent term uses the magnitude of the reading, then applies a symmetric limit around the signed indication. Use the applicable range specification.

Link to this answer
Can I calculate current or resistance limits with the same formula?

The percent-plus-counts arithmetic applies when that is how the relevant function is specified. Keep the reading and display step in one consistent unit.

Link to this answer
Why does a small reading still have a noticeable error interval?

The count term does not shrink in proportion to the indication. On a low reading it can dominate the percentage term.

Link to this answer
Does the calculated interval include reference uncertainty and probe effects?

It models the entered instrument specification only. Reference uncertainty, loading, contacts, temperature conditions and other contributions need separate consideration.

Link to this answer
Can I use a fixed-range step after the meter autoranges?

Use the step and accuracy of the range that actually produced the reading. Recalculate if autoranging changes either one.

Link to this answer
Can the error calculator prove that my particular meter is calibrated?

The output applies a published limit to an indication. It does not compare your instrument with a documented reference or establish calibration status.

Link to this answer