Meter & Probe worked guide
Build an error budget for a small shunt voltage
A small shunt protects the supply voltage but creates a tiny signal. At 0.20 A, a 0.10 Ω shunt develops only 20 mV, so voltage offset and contact placement can become important.
Content reviewed · Examples are calculations, not tests of a supplied instrument.
Estimate the signal and disturbance together
For an illustrative isolated low-voltage bench circuit, suppose 0.20 A through a nominal 0.10 Ω shunt. V = I × R gives 20 mV. P = I² × R gives 4 mW at that steady current. At 1 A the same shunt would drop 0.10 V and dissipate 0.10 W. These are calculations, not a resistor selection or a connection procedure.
Compare the calculated drop with what the circuit can tolerate at normal, startup and maximum current. Check the component’s rating and temperature behaviour; sense connections and layout can change the measured result. Do not apply this example to installed wiring or high-energy batteries.
A 50 µV offset across a 0.10 Ω shunt corresponds to 0.50 mA. That may be minor during a large active event and overwhelming beside a 20 µA sleep current. Compare error in current units, not just microvolts.

Convert voltage error into current error
Suppose the 20 mV measurement has a ±0.5 mV contribution and the shunt is 0.10 Ω ±1%. The voltage contribution alone is ±5 mA at nominal resistance. Using both bounds with I = V ÷ R gives about 193–207 mA before temperature, contact and other errors. A 1 mV display step would represent 10 mA per step; do not confuse that with an accuracy specification.
If the uncertainty or load drop is too large, an appropriately specified shunt and meter combination, or another current measurement method, is needed.
Compare with a series ammeter before choosing
The burden-voltage guide explains why the method can change the circuit. Include both the signal margin and the permitted disturbance in your selection.

Sources and review
Reviewed . Numerical examples are calculations or illustrations.